External Rampdown : External Rampdown while in Persistent Mode

a preset level. When appropriately connected to the Model 430

Programmer, these contacts can signal the Model 430 to safely and

automatically ramp the magnet field to zero, thereby preventing a magnet

quench due to low helium level in the system.

3.16.1 External Rampdown while in Persistent Mode

The following steps and associated screen displays describe the process

that occurs after external rampdown is initiated while the magnet is in

persistent mode:

1. The Model 430 Programmer first ramps the power supply to the

magnet current.

2. Once the power supply is at the magnet current, the

FIELD AT

TARGET

LED will light and the unit will momentarily “hold”:

3. Following a short “hold”, the persistent switch will be heated to

place the power supply in control of magnet current:

4. After heating (turning off) the persistent switch, the

MAGNET IN

PERSISTENT MODE

LED extinguishes, and rampdown begins.

The

FIELD AT TARGET

LED extinguishes.

5. After rampdown, the following will display:

+3.92 A

 Mode: Ramping

+0.17 Vs

PSwitch Heater: OFF

+50.00 A -

Mode: Holding

+2.11 Vs

PSwitch Heater: OFF

+50.00 A

Mode: Heating Switch (4)

+2.11 Vs

PSwitch Heater: ON

+48.85 A

 Mode: Zeroing Current

+1.89 Vs

PSwitch Heater: ON

+0.00 A -

Ext. Rampdown completed

-0.00 Vs

Press ENTER to continue