,.$ :HUNH 5(7 FRQWUROYLVF

The RET control-visc is a magnetic stirrer with heating function and viscosity trend recognition. The

instrument is suitable for precise temperature control (0 to 300 °C) of substances in containers

placed on the hot plate. The integrated stirring drive enables simultaneous stirring of the

substances with a magnetic rod in the container (0 to 1,100 rpm). The mixing intensity is dependent

on the motor speed and the size of the magnetic rod.

In the parameter window of the IKA Labortechnik RET control-visc, the desired controlling variables

and the readable measured values can be activated. The corresponding input and output paths are

marked with number fields. In the basic setting the output path 1 is set for reading out the rated

temperature.

Control

1. Input path for rated temperature.

The rated temperature is based on the medium when the temperature measuring sensor

PT 100 is inserted. If no temperature sensor is inserted the rated temperature is based on

the hot plate.

2. Input path for rated speed.
3. Viscosity trend measurement on/off.

The viscosity trend display is reset.

4. Measured values
1. Rated temperature in °C.
2. Rated speed in rpm.
3. Real temperature of measuring sensor PT 100 in °C.
4. Real temperature of hot plate in °C.
5. Safety temperature in °C.

If the rated temperature specification is greater than the safety temperature, the rated

value is ignored.

6. Real speed in rpm.
7. Viscosity trend, relative in % (corresponds to the change in the motor power

consumption).

1RWH

Please see the separate operating instructions for a detailed description of the characteristics of

the magnetic stirrer.