Hoorzolqh 6kdnhu – IKA labworldsoft User Manual

Page 153

Advertising
background image

- 153 -

<HOORZOLQH 6KDNHU

Family of processor-controlled horizontal shakers with different maximum loading and speed ranges

of 30 - 800 1/min.

&RQWURO

1. Input path for set speed.

0HDVXUHG YDOXHV

1. Set speed in 1/min.
2. Actual speed in 1/min.

0RQLWRULQJ WKH VSHHG FRQWURO

1. Watchdog active/ not active
2. Time-out time for the watchdog
3. Time-out actions.

As an additional piece of safety equipment, laboratory shakers from IKA offer device-internal

monitoring with labworld

VRIW

®

or from the attached computer. If the control software or computer

fails (crash, loss of power, etc.), this system is able to turn off the shaker or to switch it over to a

safe speed range. The shaker’s Parameters dialog box is used to activate the watchdog, to set

the time window and to set actions to be initiated in the event of a fault.

The Time-out time of the watchdog must be selected so that is greater than the current sampling

rate of the measurement by a factor of at least 5.
The features for safety monitoring are not available until after the control path for the target speed

has been activated.

127(

Detailed information on the shaker features may be found in the separate operating manual.

For proper communication the transmission parameters of the unit serial interface have to be

set to the following values:

1R VSHFLDO VHWWLQJV DUH UHTXLUHG IRU WKLV GHYLFH

Advertising