0 &rqwuro %orfn – IKA labworldsoft User Manual

Page 46

Advertising
background image

- 46 -

3:0 &RQWURO %ORFN

The

3:0

control block is a pulse spacing modulator with which data values can be digitized

IO-

Modul

) and displayed.

A data value of 0 - 100 available at the input is converted to a frequency (

mit TTL-Pegel

), the

on/off time ratio of which is the percentage equivalent of the data value.

The

IUHTXHQF\

of the output signal is given in the parameter window and can lie in the range of

0.01 to 0.1 Hz.

([DPSOHV

If a data value of



is present at the input then the output value will also be



.

If a data value of



is provided to the input then the output value will be a rectangular pulse

signal with the defaulted

IUHTXHQF\

and an On / Off ratio of 1:3 (

 RI WKH 7LPH 2Q

, 75% of

the time off).

If a data value of



is provided to the input then the output value will be a rectangular pulse

signal with the defaulted

IUHTXHQF\

and an On / Off ratio of 1:1 (

 RI WKH 7LPH 2Q

, 50% of

the time off).

If a data value of



is provided to the input then the output value will be a rectangular pulse

signal with the defaulted

IUHTXHQF\

and an On / Off ratio of 3:1 (

 RI WKH 7LPH 2Q

, 25% of

the time off).

If a data value of



is present at the input then the output value will be

TTL High

.

The

3:0

control block is primarily used in order to control devices via the digital outputs (

IKA

Laboratory Technology IO Module

) such as heating plates, heating mushrooms and others with

the help of the

control block "PID Controller"

.

In this fashion virtually any heating device can be equipped with a modern, precision temperature

control.

Carry out the following instructions:

1. Call up the parameters window by double clicking on the icon.

2. If necessary, enter the

1DPH

of the module into the "Designation" text field.

3. Select

)UHTXHQF\

.

4. Confirm entries with

2.

.