3. Time-out actions

The

5(7 FRQWUROYLVF

VDIHW\ FRQWURO

magnetic stirrer by IKA Werke offers internal monitoring by

labworldsoft

®

or the connected computer as an additional safety feature. If the control software

or the computer fails (crash/power failure etc.) this can switch off the heating and stirring

function or switch to a safe temperature and speed. You can activate the watchdog, specify the

time slot and set the action to be triggered in the event of a fault in the shaker’s parameters pop-

up window.

The watchdog time-out must be at least 20 times greater than the active sample rate of the

measurement.

The safety monitoring functions are only available once the target speed control path has been

activated.