General program functions

2.2 Login/Password

protection

2.2.1 General

Login in tiamo

tiamo can be configured so that all users must log in with User name and Pass-
word
; these entries are then checked automatically. A requirement for this is that
a User administration has been set up and the corresponding Security settings

have been made. This data is stored in the configuration database. In client/server
systems this is found on the server and applies globally for all clients (central user
administration).

FDA-conform settings

If FDA-conform work is to be carried out then the settings on the tab
Login/Password protection in the dialog window

Security settings

as per 21 CFR

11 must be switched on with the button

[Set]

. The following conditions are then

observed:

• Each time that the program is started a Login with user name and

password is required.

Password administration takes place in tiamo.
User names must be unambiguous. Once users have been entered

they cannot be deleted.

Passwords must be unambiguous for each user. No password that

has been used once by the user and has expired may be reused.

• Passwords must have a minimum number of characters.
• Passwords must have a defined Validity period after which they must

be changed.

• The number of incorrect attempts for entering the password is lim-

ited. If this number is exceeded then the user is automatically given
the status

inactive

.

Actions

If login is switched on then the following actions can be carried out:

• Login at program start
• Manual logout
• Automatic logout
• Change password