Conductometer – Metrohm tiamo 1.0 Manual User Manual

Page 606

Advertising
background image

582

Configuration

Comment on GLP test

1000 characters, [ empty ]

Comments about GLP test.

Monitoring of GLP validity

on, [ off ]

If this option is switched on then the time interval for the GLP test will be
monitored.

Note

This option can only be switched on when a date has been entered under

GLP

test date

.

GLP test interval

0 ... [ 999 ] days

Time interval to next GLP test. If an interval is entered here then the date for
the

Next GLP test

will be adapted automatically.

Next GLP test

Date selection, [GLP test date + 999 days ]

Date on which the next GLP test is to be carried out. This date can be se-
lected after pressing

in the dialog window

Select date

(see General

program functions - Edit - Select date). After the date has been entered the
value for the

GLP test interval

will be adapted automatically.

Action

If the time interval for the GLP test has expired then one of the following actions

will be triggered at both the Start test and the command sequence:

Record message

The message that the GLP test interval has expired will be saved automati-
cally in the determination.

Display message

A message appears and you can then choose whether to continue with the
run or to terminate it. If the run is continued then the message that the GLP

test interval has expired will be saved automatically in the determination.

Stop determination

The running determination will be automatically terminated. The following
message must be confirmed with

[OK]

.


Conductometer

The parameters for the 712 Conductometer are set on the following tabs:

General

General device information such as device name, device type, serial

number, etc.

RS232

Selection of the serial interface to which the Coulometer is connected.

GLP

Information about GLP test and GLP monitoring.