Commands

453

Temperature

Temperature

-20.0 ... [ 25.0 ] ... 150 °C

Manually entered measuring temperature. If a temperature sensor is con-

nected and

temperature measurement

on the tab General/Hardware is set to

automatic

or

continuous

, the temperature is measured continuously. The

value is used for the temperature adjustment during the measurement.

CAL LOOP Conc

Loop command for calibrating ion-selective electrodes. The command is used
together with the command

CAL MEAS Conc

.

Appearance

The command has the following appearance:

Parameters

The parameters for the command

CAL LOOP Conc

are configured in the following

dialog window:

CAL LOOP Conc

CAL LOOP Conc - properties

The number of standards and their concentration are defined in the dialog window

CAL LOOP Conc

.

Command name

25 characters, [ Calibration loop Conc # ]

Name of the command.

Standards

Number of standards

1 ... [ 2 ] ... 5

Selection of the number of standards for the concentration calibration.

Unit conc.

mol/L, mmol/L, [ ppm ], %, g/L, mg/L, mg/mL, µg/L

Selection of the concentration unit of the standards.

Conc. standard #

0.001 ... 999'999.999

Entry of the corresponding concentration of each standard.