Subwindow Devices

579

Note

This option can only be switched on when a date has been entered under

GLP

test date

.

GLP test interval

0 ... [ 999 ] days

Time interval to next GLP test. If an interval is entered here then the date for
the

Next GLP test

will be adapted automatically.

Next GLP test

Date selection, [GLP test date + 999 days ]

Date on which the next GLP test is to be carried out. This date can be se-
lected after pressing

in the dialog window

Select date

(see General

program functions - Edit - Select date). After the date has been entered the
value for the

GLP test interval

will be adapted automatically.

Action

If the time interval for the GLP test has expired then one of the following actions

will be triggered at both the Start test and the command sequence:

Record message

The message that the GLP test interval has expired will be saved automati-
cally in the determination.

Display message

A message appears and you can then choose whether to continue with the

run or to terminate it. If the run is continued then the message that the GLP
test interval has expired will be saved automatically in the determination.

Stop determination

The running determination will be automatically terminated. The following
message must be confirmed with

[OK]

.


Coulometer

The parameters for the 756 Coulometer and 831 Coulometer are set on the fol-

lowing tabs:

General

General device information such as device name, device type, serial
number, etc.

RS232

Selection of the serial interface to which the Coulometer is connected.

GLP

Information about GLP test and GLP monitoring.