Subwindow Sensors

645

6.7 Subwindow

Sensors

6.7.1 General

Subwindow sensors

The subwindow

Sensors

contains the Sensor table with all the configured sensors.

It can be shown in a separate window on the program part Configuration as a

part of the Configuration view or (if not present on the desktop) with

View, Quick

access

. The subwindow can be enlarged and diminished as required; it can also

be maximized.

Standard sensors

The sensor table contains the following sensors as standard; these cannot be de-
leted:

Conductivity sensor

ISE electrode

(ion-selective electrode)

Metal electrode

pH electrode

Temperature sensor

6.7.2 Sensor

table

Contents

In the sensor table the following information about the configured sensors is

shown as standard:

Sensor name

Name of the sensor.

Sensor type

Type of the sensor.

Set to work

Date on which the sensor was used for the first time.

Expiry date

Expiry date of the sensor. If sensor monitoring is switched on and if the set
date is before the current date (i.e. the working life has elapsed) then the
date will be shown in red.

Slope

Slope of the sensor in % or mV.

pH(0)/E(0)

Electrode zero point of the sensor.

Next calibration

Date on which the next calibration is to be carried out. If calibration data
monitoring is switched on and if the set date is before the current date (i.e.