Method

under the parameter

Target

:

Move

:

Special beaker

Number

:

= ### + 'Command name.LCO'

(entry with Formula editor, i.e.

= 5 + 'Calibration loop pH.LCO'

for the above sample method with the

first calibration solution on special beaker position

6

).

Defining rack positions for calibration solutions in the sample

data
A

Sample position

for the first calibration solution has to be defined in

the run window. The other calibration solutions must be placed on the
directly following rack positions. The following settings have to be
done in the command

MOVE

under the parameter

Target

:

Move

:

Rack position

Number

:

= 'MV.Sample position' – 1 + 'Command name.LCO'

(entry with

Formula editor, i.e.

= 'MV.Sample position' – 1 + 'Calibration loop

pH.LCO'

for the above sample method).

The variable

'Command name.LCO'

(index number of the calibration loop) is in-

creased by

+1

after each pass of the loop.

Note
The Formula editor is opened with a right click in the input field.

CAL LOOP pH

Loop command for calibrating pH electrodes. The command is used together
with the command

CAL MEAS pH

.

Appearance

The command has the following appearance:

Parameters

The parameters for the command

CAL LOOP pH

are configured in the following dia-

log window:

CAL LOOP pH