7 remote functions – Metrohm tiamo 1.0 Manual User Manual

Page 93

Advertising
background image

Manual control

69

2.6.7 Remote

functions

If in the subwindow for Select device a remote box or remote interface is selected
then in the subwindow for Functions/Parameters the functions and parameters
that are possible for the device will appear.

Devices

The remote functions can be carried out with the following instruments:

Titrino: 751, 758, 784, 785, 795, 798, 799

Titrando: 808, 809, 835, 836, 841, 846

Coulometer: 756, 831
Sample Changer: 730, 774, 778, 789, 814, 815

Inputs

Current status

Shows the current status of the 8 input lines.

Templates/Entry

If the current status corresponds to one of the defined Templates for the in-

puts then the corresponding name will be shown here.

Outputs

Current status

Shows the current status of the 14 output lines.

Templates/Entry

Binary pattern consisting of exactly 14 characters (0, 1, *, p), [
************** ], signal template

Input of the binary pattern for the output signal of exactly 14 characters or
selects a predefined Signal template.
The following characters can be entered:

0

= line inactive,

1

= line active,

*

= any line status

p

= set impulse (impulse length = 200 ms. If the output is to be an impulse

with a different length then an appropriate template must be defined for it.)
The output lines and bits are numbered from right to left:

Output 13 12 11 10 9 8 7 6 5 4 3 2 1 0
Bit

13 12 11 10 9 8 7 6 5 4 3 2 1 0