Commands

359

Upper limit

-200.0 ... [ 200.0 ] µA

(for

Window

=

Measured value

)

0.00000 ... [ 9999.99 ] mL

(for

Window

=

Volume

)

0 ... [ 999999 ] s

(for

Window

=

Time)

Value for the upper limit of the window.

Additional measured values

Screenshot and parameters: see DET pH - Additional measured values

MET

Command for Monotonic Equivalence Point Titrations (

MET

).

Principle

With this command titrations with reagent addition at constant volume increments

are carried out. Measured value acceptance is drift-controlled (equilibrium titra-
tion) or after a waiting period. Equivalence points are evaluated automatically.

Note

This command is suited for titrations with relatively high signal variations or sud-
denly occurring potential jump and for slow titrations or slowly responding e-
lectrodes.

Commands

Depending on the measured value one of the following four

MET

commands can

be chosen:

MET pH

Potentiometric pH measurement with pH electrodes (measured quan-

tity pH).

MET U

Potentiometric voltage measurement with metal electrodes (measured
quantity voltage U).

MET Ipol

Voltametric measurement with selectable polarization current (meas-
ured quantity voltage U).