Method

lected. In this case, the device name must be assigned by the user at the
start of the method.

Device type

Device types, [ 814 USB Sample Processor ]

Display or selection of the device type. If a device is selected under

Device

name

, this parameter is not editable any more. The device type belonging to

the selected device is displayed instead.
If the option

not defined

is selected as

Device name

, all device types being

able to execute the command can be selected independently from the de-
vices in the device table.

Pumps

Tower

[ 1 ], 2

Selection of the tower at the Sample Processor at which the pumps should
be controlled. Both Tower 1 and Tower 2 are always selectable even if the

Sample Processor has got only one tower.

Pump(s)

[ 1 ], 2, 1+2

Selection of the pump that is to be switched. The pump(s) can either be
built-in or externally connected. With

1+2

both pumps at the selected tower

will be switched at the same time.

Action

Switch on

[ on ], off

Switch on the pump(s).

Switch off

on, [ off ]

Switch off the pump(s).

Duration

on, [ off ]; 0 ... [ 10.0 ] ... 9999.9; [ s ], min

Switch on the pump(s) for a particular period.


STIR

Command for controlling a connected stirrer.

Devices

This command can be executed with the following devices:

Titrino: 751, 758, 784, 785, 794, 798, 799

Titrando: 808, 809, 835, 836, 841

Dosing Interface: 846
Coulometer: 756, 831

Sample Processor: 730, 778, 789, 814, 815

Appearance

The command has the following appearance: