Chapter 2, Getting started, Binding electrical check procedures – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 20: C2 quick calibration from the front panel, Properly installed load cell with no binding - 7, Configuration menu/selecting setup, Chapter 2: calibration

Advertising
background image

7

CHAPTER 2
Calibration

CHAPTER 2: CALIBRATION

Getting Started

The Hardy Filler/Dispenser can be calibrated two ways. The

first is with the Hardy C2

®

Second Generation calibration

which requires no test weights. C2

®

Calibration is one of

Hardy’s Core Technologies. We will describe the C2 Cali-
bration process in this chapter. The second calibration tech-
nique is called traditional calibration which requires certified
test weights. For detailed information about traditional cali-
bration See The HI 3010 Service Manual. It is important to
note that the procedures contained in this section either
explicitly stated or implied should be followed to guarantee
the performance of the instrument. Alternatives to the proce-
dures listed here are not recommended.

Before you can calibrate the instrument you first need to
check to see if the system is ready to be calibrated.

Binding

Step 1. Due a visual check to see if the load points have

been installed so that nothing is binding the weigh-
ing system. Make sure that nothing is draped over
the scale or vessel such as a hose, electrical cord,
tubes or other objects.

CAUTION:

BINDING

ON

A

SCALE

/

VESSEL

OR

LOAD

CELL

DOES

NOT

ALLOW

THE

LOAD

CELL

FREE

VERTICAL

MOVE

-

MENT

AND

MAY

PREVENT

THE

INSTRUMENT

FROM

RETURN

-

ING

TO

THE

ORIGINAL

ZERO

REFERENCE

POINT

.

Step 2. Check to see that the load sensor is mounted so that

100% of the load (Vessel with Contents) vertically
passes through the load cell always at the same
point. (See Fig. 2-1)

FIG. 2-1 PROPERLY INSTALLED LOAD CELL

WITH NO BINDING

Electrical Check Procedures

Step 1. Check to see that there is power to the controller.

If there is power to the controller The front
panel display should be lit.

Step 2. Check to see that all communication and power

cables are securely fastened to the connectors on
the rear panel.

C2 Quick Calibration From the Front Panel

Step 1. Press the Setup/3 button. The Configuration Menu

appears. (See Fig. 2-2)

Step 2. Press the down arrow until the cursor is in front of

CALIBRATION. (See Fig. 2-2)

Step 3. Press the Enter button. The CALIBRATION Menu

appears with the C2 Cal Type. (See Fig. 2-3)

FIG. 2-2 CONFIGURATION MENU/SELECTING

SETUP