Calibration menu/c2 cal, C2 calibration sub-menu, Entering the reference point – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 21: C2 calibration, Configuration menu/selecting setup

Advertising
background image

HI 3010 Filler/Dispenser/IBC

8

User Guide

FIG. 2-3 CALIBRATION MENU/C2 CAL

Step 4. Press the Enter button. C2 CAL Sub-menu appears.

(See Fig. 2-4)

FIG. 2-4 C2 CALIBRATION SUB-MENU

Step 5. The Load Sensor number is a read only field. It tells

you how many load sensors are connected to the
instrument.

Step 6. Press the down arrow button to move the cursor in

front of the Ref Point. (See Fig. 2-5)

a.

The Reference Point is the total weight that
is currently on the scale.

b.

If you have nothing on the scale the Ref
Point is 0. If you have 5 lbs on the scale the
Ref Point is 5.

FIG. 2-5 ENTERING THE REFERENCE POINT

Step 7. Press the Clr (Clear) button to clear the entry.
Step 8. Use the Alphanumeric keypad to enter the weight

that is currently on the scale.

Step 9. Press the down arrow button to move the cursor in

front of Do C2 Calibration. (See Fig. 2-6)

Step 10. Press the Enter button to complete the Calibration.

FIG. 2-6 C2 CALIBRATION

Step 11. A “function OK” momentarily appears on the

screen indicating the calibration was successful

A message that says “function Error” means
that the calibration was not successful.
Check the Troubleshooting Section of the
Technical Service manual for corrective
action.

Another message may occur which is: Secu-
rity Violation. This means that the User does
not have the security level required to do a
calibration.

Step 12. Press the Exit button until you return to the Standby

display.

Step 13. C2 calibration is complete.

C2 Quick Calibration From the Web Page

Step 1. On the Filler/Dispenser Home Page Click on Con-

figuration. (See Fig. 2-15) The Configuration page
appears. (See Fig. 2-16)

FIG. 2-7 CONFIGURATION MENU/SELECTING

SETUP

Advertising