Underfill /overfill alarm, Filler jog alarms, Did not jog alarm – Hardy HI 3010 Filler/Dispenser Controller User Manual

Page 37: 58 underfill alarm, 59 overfill alarm, 60 accept fill display, 61 did not jog alarm

Advertising
background image

HI 3010 Filler/Dispenser/IBC

24

User Guide

4.

If the Stop button is pressed the
instrument will go to Standby mode
and does not complete any more
requested cycles.

Underfill /Overfill Alarm

- - - - - ALARM CONDITION - - - - -

FIG. 3-58 UNDERFILL ALARM

FIG. 3-59 OVERFILL ALARM

OVERFILL ALARM, press the Clear but-
ton to clear the alarm. The Accept the Fill
display appears. (See Fig. 3-60) At this
time you can:

1. Accept the fill by pressing the START

button.

2. If your system allows, physically

remove some of the material until it
meets the target window requirements
and then press the START button.

3.

The instrument returns to the Waiting
On Display, repeats the motion check,
autozeros, autotares and/or checks the
OK to Fill processes and begins the
next fill cycle.

FIG. 3-60 ACCEPT FILL DISPLAY

UNDERFILL ALARM, press the Clear
button to clear the alarm, the Accept the
Fill display appears. At this time you can:

1.

Accept the fill by pressing the START
button.

2.

I your system allows, physically add
some of the material until it meets the
target window requirements and then
press the START button.

3.

The instrument returns to the Waiting
On Display, repeats the motion check,
autozeros, autotares and/or checks the
OK to Fill processes and begins the
next fill cycle.

If the gain in weight is NOT within the tar-
get window and the JOG function is turned
ON the Filler/Dispenser automatically
does the following:

1.

If the out of target window is an
OVERFILL the OVERFILL alarm
appears. Clear the alarm using the
same procedures for the OVERFILL
Alarm in the previous section.

2.

If the out of target window is an
UNDERFILL, the Jog ON display
appears.

Filler Jog Alarms

Did Not Jog Alarm

- - - - - ALARM CONDITION - - - - -

The instrument checks to see if the gate
valve or feeder of the vessel opens or
comes on. If the DID NOT JOG ALARM
appears on the display, the Filler/Dis-
penser is placed in a hold state. (See Fig.
3-61)

FIG. 3-61 DID NOT JOG ALARM

1.

Check to see why there was no Jog.
Correct the problem. Press on the
Clear button to clear the alarm. (See
Fig. 3-62)