Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 136

Advertising
background image

HI 2151/30WC MANUAL

8-16

F - Verify Individual
Load Cell Milli-volt
Readings

Press: ENTER

To check load

sensor

DVMTST
checks individual
load sensor output
mV readings.

Repeat for all load

sensors

mV

readings

acceptable?

If you are unable to determine the Milli-

volt readings. Go to

K Load sharing

OR

Contact Hardy Customer Support

800-821-5831

F

TEST COMPLETE

Yes

Record load
sensor mV output
level for
comparison.

Press TST/CLR

START

INTEGRATED TECHNICIAN

( I.T.)

diagnostic program

Select DVMTST

DVMMV

Press: ENTER

NEXT

Using the load cell certificate, verify the Milli volt per volt rating
Example: 3 mV/V load cells will produce approximately 15 mV at full load.
That is 5 volts excitation X 3 mV/V. A scale capacity of 1,000 lbs with 100
lbs of deadload at empty the loadpoint mV reading should equal 1.5 mV.

Zero mV

reading

High mV

reading

1) Stressed load cell, remove
all load and re-test.
2) Excessive loading.
For additional
testing go to:

NO

Unstable

reading

Defective load cell?
Replace and repeat

TEST

F

Negative

mV reading

A

1) No dead load.
Apply load and re-test.
2) Wiring error.
Verify color code using the
load cell certificate.
3) Open bridge circuit.
Disconnect power and verify
loadpoint bridge resistance
reading with an Ohmmeter.

H

K