Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 146

Advertising
background image

HI 2151/30WC MANUAL

8-26

DISPLAY

TEST

FAILURE SOLUTIONS

RAMTST

Ram Test - Writes to each location

None

in RAM & reads back and verifies data.
Test is non-destructive.

PASSED or

Indicates whether RAM test passed

1. With most RAM failures

system will not operate.

FAILED

failed.

1. Check that EPROM (U13)

on main board is seated.

2. Contact

Hardy

Instruments for assistance.

OE OE

Two hex numbers displayed; right one

Discrepancy between two

(example)

is actual checksum, left one is

numbers indicates EPROM

calculated from test. The two numbers must faulty.
match. For each revision to EPROM
numbers will be different.

PASSED or

If the two numbers matched, test will

If failed, replace EPROM.

FAILED

pass; otherwise fail is indicated. It is

(If this doesn't eliminate

possible for EPROM to loose one or more the problem, there could be a
bits & still operate.

BUS problem.)

SLOT 1

Indicates next display is for slot 1.

NONE or

Indicates which type of option card is

1. Check connector for

OPTION

installed in slot 1, if any.

proper mating.

2. Try card in another slot.

SLOT 2

Indicates next display is for slot 2.

1. Check connector for

NONE or

Indicates which type of option card is

proper mating.

OPTION

installed in slot 2, if any.

2. Try card in other slot.
3. Swap with spare option card.

CALTYP

How instrument was calibrated

None

NOTE:

One of the next three messages will be displayed, depending on the type of calibration
that was performed

C2 CAL

C2TM Second Generation Calibration

None

H CAL

Hardware Calibration

None

S CAL

Software Calibration

None

UNITS

Indicates next parameter will be units

None

of measure.