About the bcd board print output signal -6, Analog output board -6, Bcd jumper settings -7 – Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 60: W1 w2 w1 w2 positive pulse negative pulse, 6 fig. 4-6 analog output board

Advertising
background image

HI 2151/30WC MANUAL

4-6

FIG. 4-6 ANALOG OUTPUT BOARD

Parallel BCD Board Print
Configuration Procedures -
B2, -B5, -B9

About the BCD Board
Print Output Signal

Some receivers require either a positive (high) pulse or a negative (low)
when printing. Select the jumper that meets the receiver input signal
requirements.

Jumper Location

The switch is marked W1 W2 in the top center of the board.

Jumper Settings (See
Fig. 4-7 and 4-8)

Positive Pulse = W1

Negative Pulse = W2 (Default)

FIG. 4-7 BCD CARD JUMPER SETTING FOR POSITVE PULSE

FIG. 4-8 BCD JUMPER SETTING FOR NEGATIVE PULSE

0535-0406-

REV-

ANALOG OUT PWA

C1

U1

C4

RN

1

C3

C2

U2

U3

U4

C6

U6

S/N

U8

S

D

D

Q1

C5

U7

1

2

25

26

P1

U5

RN

2

C9

R1

R2

R3

C1

1

C8

CR1

C1

7

C1

8

C13

C16

C7

TP1

C1

4

C1

5

R4

R5

C19

R15

R14

R13

R12

R11

R10

R9

R8

R7

U9

R6

C20

O

I

VR2

-

+

+

+

VR3

C21

C22

C24

C23

C25

U10

R16

R19

TP2
ISO
GND

VR4

C

E

Q3

J1

1

6

D

Q

2

G

C26

1
2
3
4

R1

7

R1

8

CR

2

R

2

0

C

R

3

C

R

4

R

2

1

W

Voltage Zero

Current Span

Voltage Span

Current Zero

W

1

W2

W1

W2

Positive
Pulse

Negative
Pulse

Advertising