HI 2151/30WC MANUAL

5-12

Use left arrow

← to move cursor one space to the left. The

letter "T" starts flashing.

Use the up arrow

↑ to select "T" (to output the tare value),

or "_" (tare value will not be output).

Use left arrow

← to move the cursor one space to the left.

The letter "N" starts flashing.

Use up arrow

↑ to select "N" (to output net weight) or "_"

(will not output net weight).

Use the left arrow

← to move the cursor one space to the

left. The letter "G" starts flashing.

Use up arrow

↑ to select "G" (to output gross weight), or

"_" (will not output gross weight).

NOTE:

At least one of the letters "G", "N", "T", "R", or "S" must be selected or Error 54 will
be displayed.

Step 13. Press the Enter button twice. First the control Sub-Menu

CONTRL appears; then SOFTRE or HARDRE appears.

Step 14. Use the up or down arrows

↑↓ to select either option.

Step 15. Press the Enter button twice. First the Echo Sub-Menu ECHO

appears; then ON or OFF will be displayed).

Step 16. Use the up or down arrows

↑↓ to select either parameter. For

our example select OFF.

Step 17. Press the Enter button twice. First the Address Sub-Menu

ADDRES appears; then the current two-digit address appears.

Step 18. Press the -/Test/Clr button to clear the display.

Step 19. Enter the desired instrument address, 99 maximum. 10 (unit

10 is used for this example).

Step 20. Press the Enter button three times to return to the weight dis-

play and normal operation.

NOTE:

* Available only when "BI-DIR" mode is selected

Analog Output Setup
Procedures (Option)

About Analog
Output (See Chapter
4 for Configuration
Procedures)

The analog output option allows the user not only to output gross, net,
ROC, Peak, or total weight as 0-5V, 0-10V, 0-20 mA or 4-20 mA (or
the reverse of these via the front panel), but makes it possible to span
these ranges over a portion of the weight data. A full analog output is
obtained over the range desired. Resolution of the analog out is 16,000
counts, or the number of display counts available in the range selected,
whichever is less. All of this is accomplished via the front panel or the
serial port. Two analog option boards may be installed in each unit.