N - Blank Screen

N

Monitor system for proper

operation.

Check out complete

Weight Controller's front display is blank

1)

Check for proper power at J6
and the source connection.

2)

Check the circuit breaker at the
source

Measure the excitation

voltage at J1.

5 vdc ?

Replace the 1/2

amp slo-blow fuse

and apply power.

Does the fuse blows

again?

Check the power fuse located inside

the case.

OK?

Measure A.C. Power

at J6

OK?

Measure the

Excitation voltage

at J1.

5 vdc?

Disconnect all the

connectors from the

back panel except

power.

1) Reconnect the jacks one
at a time checking the 5 vdc
excitation.
2) If reconnecting any jack
effects the 5 vdc, check for
wiring errors.

Contact

Hardy Instruments

Customer Service

800-821-5831

No

Yes

No

No

YES

No

Yes

Yes

Contact

Hardy Instruments

Customer Service

800-821-5831

NO

Display OK

Yes