HI 2151/30WC MANUAL

4-8

Dip Switch Location
(See Fig. 4-10)

FIG. 4-10 AUTO ZERO DIP SWITCH SETTING

Dip Switch Setting

Step 1.

To enable the Auto Zero Tolerancing, set dip switch #8 to ON.

Step 2.

To disable the Auto Zero Tolerancing, set dip switch #8 to
OFF.

Configuring Peak Hold
Mode

About Configuring
Peak Hold

Highest gross or averaged gross is selectable on dip switch S2 position
number 2, located internally on the Main Controller board. (See Fig. 4-
11)

NOTE:

Dip Switches S2 on the Power Relay Board are all normally in the OFF position.
Contact Hardy Instrument Customer Support Department before changing any of
these switch positions.