Power & relay board dip switch location -10 – Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 64

Advertising
background image

HI 2151/30WC MANUAL

4-10

2. When the switch is OFF, Tare, Mode, Zero, and lb/kg keys are

available.

Configuring the Power &
Relay Board Dip Switches
S2 (See Fig. 4-12)

FIG. 4-12 POWER & RELAY BOARD DIP SWITCH LOCATION

1. When the switch is ON, incoming checksums are ignored. When

the switch is OFF incoming checksums are read.

2. S2-2 - When the switch is ON, the Peak Hold signal read is aver-

aged. When the switch is OFF, the Peak Hold signal is instanta-
neous.

3. S2-3 - When the switch is ON, the instrument is in the NBS mode

of operation. Resolution is limited to 1:10,000 counts. When the
switch is OFF, the resolution is 1:985,000.

4. S2-4 - This switch must be toggled (position changed) to enter NBS

calibration from the front panel CAL button.

5. S2-5 - Turns Off > character on print output serial port.

6. S2-6 - Puts unit in Blind Remote Mode.

7. S2-7 - is not used.

8. S2-8 - is used for resetting Calibration & Configuration to the Fac-

tory Defaults