Chapter 2 - specifica- tions – Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 21

Advertising
background image

Chapter 2 - Specifications

2-1

CHAPTER 2 - SPECIFICATIONS

Chapter Two covers the specifications for the HI 2151/30WC weight
controller and optional equipment. The specifications listed are
designed to assist in the installation, operation, and troubleshooting.

HI 2151/30WC
Specifications for a
Standard Instrument

Conversion Rate

20 updates per second

Resolution

Displayed:

1:985,000 (@ 3 mV/V)
1:656,000 (@ 2 mV/V)

Internal:

1:1,048,576

Excitation Voltage

0-5 VDC

Averages

1 to 200 User Selectable in single Increments

Input

Up to eight (8) 350 ohm Full Wheatstone Bridge, Strain Gauge
Load Sensors/Cells (5 volt excitation) on one vessel.

Signal Voltage Range0-15mV

Display

6 digit, 14 segment red LED, 0.6" alpha-numeric

Bar-Graph 30 segment LED

Display Increments
(Graduations)

1,2,5,10,20,50,100,200,500 user selectable via the front panel key
pad.

Corresponding weight is dependent on the decimal point location.

Key Pad

16 tactile keys

6 dedicated functions

10 dual numeric and function

Standard Setpoint
Relays

2 Standard SPDT (Form C) Relays

Wire Size: 12 AWG Maximum

Maximum Switch Current: 5 Amps

Maximum Switch Power: 150 Watts or 1250 Volt/Amps

Maximum Switch Voltage: 5 Amps @ 3 VDC or 125 VAC

Non-Linearity

0.0015% of Full Scale

Maximum Zero
Tolerance

32766

WAVERSAVER

®

7.5 Hz

3.5 Hz

1.0 Hz

0.5 Hz

0.25 Hz