Electrical check procedures -2, Properly installed load cell w/no binding -2 – Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 80

Advertising
background image

HI 2151/30WC MANUAL

6-2

FIG. 6-1 PROPERLY INSTALLED LOAD CELL W/NO BINDING

Step 3.

Check to see that nothing is coming in contact with the scale/
vessel other than service wires and piping that have been
properly mounted with flexible connectors.

Electrical Check
Procedures

Step 1.

Check to see that there is power to the controller.

If there is power to the controller The front panel display
should be lit.

If an error message (ERREXC) appears in the display, it
means there is a problem in the system. See Troubleshoot-
ing Chapter 8 for corrective action.

If the display appears with a value the unit is ready for cali-
bration.

Step 2.

Typical Load Cell/Point Input/Output Measurements (EXC &
SIG Outputs)

The HI 2151/30WC is designed to supply 5 vdc excitation
to as many as eight 350 ohm load cells/points.

Vessel

Direction of
the applied
load