Carrier Streamline Scroll 69NT20-531-300 User Manual

Page 55

Advertising
background image

T-309

4-3

3.

Pre-trip may also be initiated via

communications. The operation is the same

as for the key pad initiation described below

except that should a test fail, the pre-trip

mode will automatically terminate. When

initiated via communications, a test may not

be interrupted with an arrow key, but the

pre-trip mode can be terminated with the

PRE-TRIP key.

a. Press the PRE-TRIP key. This accesses a test selec-

tion menu.

b. TO RUN AN AUTOMATIC TEST: Scroll through

the selections by pressing the UP ARROW or

DOWN ARROW keys to display AUTO, AUTO 1 or

AUTO 2 as desired and then press the AUTO key.

1. The unit will execute the series of tests without any

need for direct user interface. These tests vary in

length, depending on the component under test.

2. While tests are running, “P#-#” will appear on the left

display, where the #’s indicate the test number and

sub-test. The right display will show a countdown

time in minutes and seconds, indicating how much

time there is left remaining in the test.

CAUTION

When a failure occurs during automatic

testing the unit will suspend operation

awaiting operator intervention.

When an automatic test fails, it will be repeated once .

A repeated test failure will cause “FAIL” to be shown

on the right display, with the corresponding test num-

ber to the left. The user may then press the DOWN

ARROW to repeat the test, the UP ARROW to skip to

the next test or the PRE--TRIP key to terminate test-

ing. The unit will wait indefinitely, until the user

manually enters a command.

CAUTION

When Pre--Trip test Auto 2 runs to comple-

tion without being interrupted, the unit will

terminate pre-trip and display “Auto 2”

“end.” The unit will suspend operation until

the user depresses the ENTER key!

When an Auto 1 runs to completion without a failure,

the unit will exit the pre-trip mode, and return to nor-

mal control operation. However, dehumidification

and bulb mode must be reactivated manually if re-

quired.

c. TO RUN AN INDIVIDUAL TEST: Scroll through

the selections by pressing the UP ARROW or

DOWN ARROW keys to display an individual test

code. Pressing ENTER when the desired test code is

displayed.

1. Individually selected tests, other than the LED/Dis-

play test, will perform the operations necessary to

verify the operation of the component. At the con-

clusion, PASS or FAIL will be displayed. This mes-

sage will remain displayed for up to three minutes,

during which time a user may select another test. If

the three minute time period expires, the unit will

terminate pre-trip and return to control mode opera-

tion.

2. While the tests are being executed, the user may ter-

minate the pre-trip diagnostics by pressing and hold-

ing the PRE-TRIP key. The unit will then resume

normal operation. If the user decides to terminate a

test but remain at the test selection menu, the user

may press the UP ARROW key. When this is done

all test outputs will be de-energized and the test

selection menu will be displayed.

3. Throughout the duration of any pre-trip test except

the P-7 high pressure switch tests, the current and

pressure limiting processes are active .

d. Pre-Trip Test Results
At the end of the pre-trip test selection menu, the

message “P,” “rSLts” (pre--trip results) will be

displayed. Pressing the ENTER key will allow the user

to see the results for all subtests (i.e., 1-0, 1-1, etc). The

results will be displayed as “PASS” or “FAIL” for all the

tests run to completion since power up. If a test has not

been run since power up, “

-- -- -- -- --

” will be displayed.

Once

all

pre--test

activity

is

completed,

dehumidification and bulb mode must be reactivated

manually if required.

4.8 OBSERVE UNIT OPERATION
4.8.1 Probe Check

If the DataCORDER is off or in alarm the controller will

revert to a four probe configuration which includes the

DataCORDER supply and return air probes as the

secondary

controller

probes.

The

controller

continuously performs probe diagnosis testing which

compares the four probes. If the probe diagnosis result

indicates a probe problem exists, the controller will

perform a probe check to identify the probe(s) in error.
a. Probe Diagnostic Logic -- Standard
If the probe check option (controller configuration code

CnF31) is configured for standard, the criteria used for

comparison between the primary and secondary control

probes is:

1_C (1.8_F) for perishable set points or 2_C (3.6_F)

for frozen set points.
If 25 or more of 30 readings taken within a 30 minute

period are outside of the limit, then a defrost is initi-

ated and a probe check is performed.

In this configuration, a probe check will be run as a part

of every normal (time initiated) defrost.
b. Probe Diagnostic Logic -- Special
If the probe check option is configured for special the

above criteria are applicable. A defrost with probe check

will be initiated if 25 of 30 readings or 10 consecutive

readings are outside of the limits
In this configuration, a probe check will not be run as a

part of a normal defrost, but only as a part of a defrost

initiated due to a diagnostic reading outside of the

limits.
c.The 30 minute timer will be reset at each of the follow-

ing conditions:

Advertising