Indications of the type plate

Information

1,20 kW

Nominal power 1.20 kW

230 V 1 N ~

Nominal voltage 230 V ± 10%, single-phase unit

5,3 A

Nominal current 5.3 Amp

50/60 Hz

Power frequency 50/60 Hz

Explosion proof inner chamber

Explosion proof inner chamber

Ex classification acc. to 94/9/EC

Ex classification according to ATEX Directive 94/9/EC

II -3G IIB T3-T1 Gc X

Ex classification of the inner chamber:

II -3G IIB T3-T1 Gc X

Max. temp. of inner surface 320 °C

Maximum temperature of the inner surfaces: 320 °C / 608°F

Temp. class depending on setting of

temp. safety device

Temperature class acc. to IEC 60079-14 for entire unit: T1,

T2 or T3 depending on the setting of the temperature safety

device

Max. operating pressure 125 +/- 25

mbar

Max. operating pressure (heating release): 125 +/- 25 mbar


Symbol

Information

CE conformity marking

Ex classification acc. to ATEX Directive 94/9/EC, referring to the inner chamber

TÜV mark of conformity (Mark P)

Electrical and electronic equipment manufactured / placed on the market in the EC

after 13 August 2005 and to be disposed of in a separate collection according to di-

rective 2002/96/EC on waste electrical and electronic equipment (WEEE).

The equipment is certified in the GOST R certification system of GOSTSTANDARD

Russia.


1.5 General safety instructions on installing and operating the vacuum drying

oven

With regard to operating the vacuum drying oven VDL and to the installation location, please observe the

guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120 or ZH 1/119 labora-

tory guidelines issued by the employers’ liability insurance association) (for Germany).
BINDER GmbH is only responsible for the safety features of the unit provided skilled electricians or quali-

fied personnel authorized by BINDER perform all maintenance and repair, and if components relating to

chamber safety are replaced in the event of failure with original spare parts.
To operate the unit, use only original BINDER accessories or accessories from third-party suppliers au-

thorized by BINDER. The user is responsible for any risk caused by using unauthorized accessories.