2 programming with setting “step – BINDER VD 23 User Manual

Page 57

Advertising
background image

VDL (E2.1) 04/2014

page 57/107

8.1.2 Programming with setting “step”

With setting “Step” selected, you don’t need to program the transition section in the Program Editor.

With setting “step” the controller will equilibrate only to constant temperatures; programming

ramps is no longer possible.

The set-points are maintained constant for the duration of a program section. At the start of each program

section, the unit heats up with the maximum speed in order to attain the entered set-point

Program entry as set-point step (example):

W/°C

t/min.

S01 S03 S05 S07

0

20

40

60

80

100

30

120

180

380

Program table corresponding to the diagram (with setting “Step”):

Section


SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

S01

40

00:30

S02

60

01:30

S03

80

01:00

S04

20

03:20

You can now enter the values of such a program table to the RD3 program controller (chap. 8.2).