VDL (E2.1) 04/2014

page 68/107


Time of the real time clock (Time)
Main menu. Use the program key to access the settings of hour and minute in the corresponding sub-

menus.

There is no automatic switch between daylight saving time and regular time.

Hour of the real time clock (Hour)
Enter the hour (0 up to 23).

Minute of the real time clock (Minute)
Enter the minute (0 up to 59).

Normal Display

Display 1 shows

e.g. 39.8

C

(actual temperature value)

Display 2 shows e.g. 15.05.06 13:52 - - (actual date and time, actual switching state of week pro-

gram timer channel 1: Off, channel 2: Off)

Press down key

for 5 sec

Display 1 shows

e.g. 0000

Display 2 shows PROGRAM EDITOR

(you are in the program editor)

Press down key

for 5 sec

Display 1 shows

0000

Menu visible only if week program timer is activated.

Display 2 shows WEEK PROG. EDITOR

(you are in the week program editor)

Press down key

for 5 sec

Display 1 shows

0000

Display 2 shows

USER – LEVEL

(you are in the user level)

Press program key

Display 1 shows

0000

Display 2 shows UserCod?

0000

(enter user code, display flashes)

Enter the user code using arrow keys

e.g.

0001 (basic setting,

or the valid code in case it has been previously

changed in this menu).

Value is shown in both displays.

Automatically forward after 2 sec

Display 1 shows

1

(actual address: 1)

Display 2 shows

Adress

1

(entry of unit address)

(actual address: 1)

Enter the unit address (1 up to 254) using

arrow keys

Address is shown in both displays.

Press key

Display 1 shows

1

(actually valid user code: 1)

Display 2 shows User-cod

1

(change user code)

(actually set: 1)

Enter a new value using arrow keys

Value is shown in both displays.