WARNING

High concentration of inert gas.
Risk of death by suffocation.

Do NOT set up units in non-ventilated recesses.

Ensure technical ventilation measures.

Respect the relevant regulations for handling these gases.

When decommissioning the vacuum drying oven, shut off the inert gas supply: Close

valve (6).


4.

Installation and connections

4.1 Vacuum expansion racks


The low-loss heat transfer to the material occurs via the pa-

tented aluminum vacuum expansion racks (also available in

stainless steel as an option). The strong tension causes the

racks to fit tightly against the interior wall and their large-

surface contact area ensures rapid and effective heat transfer.
The removable rack holders allow for easy positioning.
You can also remove the expansion racks for easy cleaning.

Do not remove them too often in order to prevent wear.

Figure 12: Using the expansion racks

Pushing the locking lever: The expansion rack is released and can be removed.

Pulling the locking lever: The expansion rack is pressed against the inner chamber walls.

Following each new tightening of an expansion rack, check that the lateral parts of the rack fit

closely over their whole surface to the inner chamber wall. This is necessary in order to ensure

the specified temperature exactitude.

CAUTION

Invalid calibration.

Do NOT change between aluminum and stainless steel racks

Use the delivered expansion racks only