Troubleshooting, Caution – BINDER VD 23 User Manual

Page 93

Advertising
background image

VDL (E2.1) 04/2014

page 93/107

17.5 Disposal of the unit in non-member states of the EC

CAUTION

Alteration of the environment.

For final decommissioning and disposal of the vacuum drying oven, please contact

BINDER Service.

Follow the statutory regulations for appropriate, environmentally friendly disposal.

The main board of the vacuum drying oven includes a lithium cell. Please dispose of it according to na-

tional regulations.

18. Troubleshooting

Fault description

Possible cause

Required measures

General

Unit permanently turned off.

No power supply.

Check connection to power supply.

Wrong voltage.

Check power supply for voltage of

115V or 230V.

Oven not connected to com-

pressed air supply.

Connect the oven to the com-

pressed air supply before start-up.

Unit fuse has responded.

Check unit fuse.

Safety device has turned off the

oven.

Allow the oven to cool down and

press the “RESET” button. Check

temperature set-point and setting of

safety device (chap. 12). If appro-

priate, select suitable limit value.

Safety device defective.

Contact BINDER Service.

Controller defective.

Heating
Chamber doesn’t heat up.
LED (7a) “Heating active” lit.
Pilot light (4) “No heating re-

lease” lit.

Pressure > 125 ± 25 mbar / 3.69

±0.74 inHg.

A vacuum of 125 ± 25 mbar / 3.69

±0.74 inHg.) or below is needed for

heating release.

Unit door not properly closed.

Completely close unit door.

Door gasket defective.

Replace door gasket.

Chamber doesn’t heat up.
LED (7a) “Heating active” lit.
Pilot light (4) “No heating re-

lease” not lit.

Heating element defective.

Contact BINDER Service.

Semiconductor relay defective.

Chamber doesn’t heat up.
LED (7a) “Heating active” not lit.

Safety device has turned off the

oven.

Allow the oven to cool down and

press down the “RESET” button.

Check temperature set-point and

setting of safety device (chap. 12).

If appropriate, select suitable limit

value.

Controller defective

Contact BINDER Service

Set-point temperature is not

reached after specified time.

Controller not adjusted.

Calibrate and adjust controller.