Next step –set-point entry in the desired program sections:
Basic entry principle:
Access the parameters of individual program sections with button X/W one after

the other. Enter the values of the individual parameters with the arrow keys. A value flashing once after 2

seconds indicates that it has been adopted by the controller. If several parameters are to be skipped (e.g.

in order to change a parameter in a posterior program section), the parameters can be rapidly jumped

over by holding down the X/W key. If no button is pressed for more than 120 sec the controller switches

back to Normal Display. The program entered to this point remains stored.

Press key

Display 1 shows

e.g. 40.0

C

(actual temperature set-point)

Display 2 shows

alternating

S01: TEMP 40.0

(variable: temperature in °C)

CONTINUE X/W

(information: go on with X/W)

Enter temperature set-point of S01 in °C

using arrow keys

Press key

Display 1 shows

e.g. 00.30

(actual section length set-point)

Display 2 shows

alternating

S01: TIME 00:30

(variable: section length in hh:mm)

CONTINUE X/W

(information: go on with X/W)

Enter section length set-point of S01 in

hh.mm using arrow keys

Value is shown in both displays.

Press key

Selecting the next program sections to be entered

Display 1 shows

e.g. 02

(section S02 selected)

Display 2 shows

alternating

P01: --- SEC.

Section S02 has already been created.

Enter new set-points for the individual parameters with X/W.

CONTINUE X/W

or:

Display 1 shows

e.g. 02

(section S02 selected)

Display 2 shows

alternating

P01: --- SEC.

Section S02 has not yet been created.

Enter set-points for the individual parameters with X/W

NEW SEC. X/W

Select the next section to be entered using

arrow keys

Display 1 shows

e.g. 60.0

C

(actual temperature set-point)

Display 2 shows

alternating

S02:TEMP 60.0

(variable: temperature in °C)

CONTINUE X/W

(information: go on with X/W)

Enter the temperature set-point of S02 in °C

using arrow keys

Etc.

If all sections up to S10 or up to S20 have been programmed, section S01 follows again. In order to quit

the entry mode, press the “

EXIT” button several times or wait 120 sec. → the controller will then return to

Normal Display.

When changing the set-point, check the setting of the safety device (chap. 12).