Program type selection (PrgSelec)
Select between entry of two programs with up to 10 sections each or of one program with up to 20 sec-

tions.

When changing from 2 programs to 1 program or vice-versa, existing programs are de-

leted in the program editor!

Maximum section duration (Prg.Time)
The maximum length of an individual program section can be set to either 99 h 59 min or to 999 h 59

min. This setting is then valid for all program sections.

When changing the maximum duration setting, pre-existing programs will be deleted in

the program editor.

Set-point programming type (Setp.sim)
Selection between “Ramp” and “Step”. With setting “Step” selected, you don’t need to program the

transition section in the Program Editor.

If you select setting “step”, the controller will equilibrate only to constant temperatures;

programming ramps becomes impossible.

A change between settings “ramp” and “step” will influence all programs. Note that signifi-

cant change in time courses may arise in existing programs.

Tolerance limit range (Tol.band)
Entry of a tolerance limit value in °C. If the actual value of temperature exceeds the set-point of a pro-

gram section by more than the entered tolerance limit value, the program is halted (LED (7d) flashes)

until the actual temperature value is again within the tolerance range
Entry of “0” means tolerance limits are off.

Activating or inactivating the week program timer (Prog.Clk)

Inactive”: The week program timer is turned off (factory setting). The corresponding setting menu

(chap. 7) is not visible, nor is set-point 2 in the “Fixed value entry mode” (chap. 6).

Active”: The week program timer is activated.

When deactivating the week program timer, any programming made in advance will re-

main in memory and take effect when the week program timer is activated again.

Deactivate the week program timer before staring a program (chap. 9).

Display mode (12h/24h)
Select between 12 hours (display “AM” or “PM”) or 24 hours.

Date of the real time clock (Date)
Main menu. Use the program key to access the settings of year, month, and day in the corresponding

submenus.

Year of the real time clock (Year)
Enter the year (2006 up to 2050)

Month of the real time clock (Month)
Enter the month (1 up to 12).

Day of the real time clock (Day)
Enter the day (1 up to 31).