The object temperature is indicated on Display 2 of the RD3 controller.

Display 1 shows

e.g. 24.6

C

(actual temperature value)

Display 2 shows

24.4 15:41

(actual object temperature in °C, actual time)

Figure 23:

Standard Display without the

object temperature display option

Figure 24:

Display with the

object temperature display option

The object temperature data is put out combined with the temperature data of the controller to the RS 422

interface as a second measuring channel. This permits recording by the BINDER documentation software

APT-COM™ DataControlSystem (option, chap. 15.6).

Assembly and connection of the object temperature recording

Insert the Pt 100 temperature sensor from the rear through the measuring connection (15) into the

inner chamber.

The 3 contacts of the Pt 100 sensor are conducted outside via a measuring access port. From there,

establish the connection to the DIN socket (14) at the top of the rear panel of the unit marked “Pt100”.

For reasons of explosion protection, this electrical connection to the inner chamber is conducted via a

triple internal safety barrier with a conducting-state voltage of 1.6 Volt maximum against ground.

Figure 25:

Figure 26:

Figure 27:

Measuring connection (15)

with measuring access port

Plug for connection cable to

DIN socket “Pt 100”

Measuring current port with universal

eccentric ring and locking ring DN 16, with

flexible Pt 100 temperature sensor

Pt100 connection (14) at

the rear of the unit


Technical data of the Pt 100 sensor:

Three-wire technique

Class B (EN 60751)

Temperature range up to 300 °C / 572°F

Stainless steel protective tube, length 45 mm / 1.77 in, stainless steel material no. 1.4501

If the oven is charged to full capacity, depending on the load, deviations from the specified