BINDER VD 23 User Manual

Page 64

Advertising
background image

VDL (E2.1) 04/2014

page 64/107


In the first step, select a program. This is on condition that a program has been entered previously (chap.

8.2) and that program type “2 programs with 10 sections each” has been selected in the user level (chap.

10).
Then define the settings for the program course. Two parameters can be set:

Program delay time, i.e. a defined time before a program starts. It can be entered with a precision of 1

minute, and its maximum value is 99.59 (99 h 59 min). If the value is 00.00, the program will start im-

mediately.

Number of program cycles, i.e. the desired number of program repeats. Values from 1 to 99 can be

entered. If the program is not going to be repeated, enter the value “0”. For indefinite repeats enter the

value “–1”. The program is repeated as a whole; it is not possible to repeat individual sections.

In the last step start the selected program. These steps must be carried out subsequently.

Deactivate the week program timer (factory setting, setting in the user level, chap. 10) before

starting a program.



Step 1 – Program selection (only with program type “2 programs” set)

Normal Display

Press program key

Display 1 shows

e.g. 1

(actual selection of the program)

Display 2 shows

SEL.PRG.

(select program 1 or 2)

Enter program number 1 or 2 with

arrow keys

Value is shown in display 1.



Next step – entry of program course settings

Press program key

Display 1 shows

e.g. 00.00

(entered delay time hh.mm)

Display 2 shows

RUN TIME

(enter delay time of program start)

Set delay time in hh.mm using arrow keys

Value is shown in display 1.

Press program key

Display 1 shows

e.g. -1

(actual selection of the number of program cycles)

Display 2 shows

REPEAT

(enter number of program cycles)

Select number of cycles -1, 0, 1 etc. using

arrow keys

Value is shown in display 1.

Advertising
This manual is related to the following products: