BINDER VD 23 User Manual

Page 59

Advertising
background image

VDL (E2.1) 04/2014

page 59/107

You can now enter the values of the program table to one of the program places of the RD3 program con-

troller.

Step 1 – Selecting the program and the first program section to be entered:

Normal Display

Display 1 shows

e.g. 39.8

C

(actual temperature value)

Display 2 shows e.g. 15.05.06 13:52 - -

(actual date and time, actual switching state of week pro-

gram timer channel 1: Off, channel 2: Off)

Press down key

for 5 sec.

Display 1 shows

e.g. 0000

Display 2 shows PROGRAM EDITOR

(you are in the program editor)

Press program key

Display 1 shows

0000

Display 2 shows

UserCod? 0000

(enter user code)

Enter user code using arrow keys

e.g.

0001

(basic setting, adjustable in the user level,

chap. 10).

Value is shown in both displays.

Automatically forward after 2 sec.

Display 1 shows

e.g. 01

(program P01 selected)

Display 2 shows

alternating

--- : --- PRG.

(program can be selected)

CONTINUE X/W

(information: to 1

st

program section with X/W)

Select program P01 or P02 using arrow keys

Value is shown in display 1.

Press key

In the selected program P01 or P02, program sections can be selected:

Display 1 shows

e.g. 01

(section S01 selected)

Display 2 shows

alternating

P01: --- SEC.

Section S01 has already been created.

Enter new set-points for the individual variables with button

X/W

CONTINUE X/W

or:

Display 1 shows

e.g. 01

(section S01 selected)

Display 2 shows

alternating

P01: --- SEC.

Section S01 has not yet been created.

Enter set-points for the individual variables with button X/W

NEW SEC. X/W

Select sections S01 to S10 or to S20 using

arrow keys

As long as no program section has been entered, the display switches back to 01 in case of any entry

> 01, because all sections need to be entered one after the other, and each new section is created as

NEWSEC.

Example: If three program sections have been already entered, the next section to be entered is S04.

Before this, no section > S04 can be selected.

Advertising