BINDER VD 23 User Manual

Page 69

Advertising
background image

VDL (E2.1) 04/2014

page 69/107

Press key

Display 1 shows

0000

(no function)

Display 2 shows

Saf.mode: Limit

(no function)

Press key

Display 1 shows

0

(no function)

Display 2 shows

Saf.setp

0

(no function)

Press key

Display 1 shows

0000

(no function)

Display 2 shows Decimal: XXX.X

(setting of decimal point position)

(actual setting: XXX.X)

Select decimal point position using arrow

keys

Decimal point position XXX.X or XXXX. is

shown in display 2.

Press key

Display 1 shows

0000

(no function)

Display 2 shows Buzzer : Active

(setting of the alarm buzzer)

(actual setting: „Active“)

Select between “Active” and “Inactiv” using

arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

0000

(no function)

Display 2 shows Language : English

(selection of controller language)

(actual setting: English)

Select between languages German, English,

and French using arrow keys

Setting is shown in display 2.

Press key

Display 1 shows

e.g. 0004

(unit operating hours)

Display 2 shows Oper.hs 0004:28

(operating hours up to now hhhh:mm)

(actually shown: 4 hs 28 min)

Press key

Display 1 shows

1000

(actual setting: 1000 hs)

Display 2 shows

Op.limit

1000:00

(maximum number of operating hours

hhhh:mm (actual setting: 1000 hs)

Set value using arrow keys

Setting is shown in both displays.

Press key

Display 1 shows

0000

(no function)

Display 2 shows Op.back : No

(reset counter of operating hours ?)

(actual setting: No)

Select between “Yes” and “No” using arrow

keys

Setting is shown in display 2.