15.2 Vacuum module empty (without pump) (option)

The mounting instructions Art. no. 7001-0137 supplied with the vacuum module describe

how to mount the VDL vacuum drying oven onto the vacuum module and installing the

suction line to the vacuum module.


(13)



Connection to a vacuum source
Connect the vacuum connection (13) (small flange DN 16) at the

rear of the unit at the top to a vacuum pump or to a domestic

vacuum system via a vacuum hose or a fixed vacuum pipe.
When using a vacuum hose, we recommend using the BINDER

connection kits for VP4 or VP5 (chap. 15.1). The module has an

appropriate hose outlet at the back.










Figure 19: VDL 53 with the option “Vacuum module”, mounted,

and installed vacuum hose

CAUTION

Technical defects due to excessive negative pressure.
Danger of implosion.
Damage to the unit.

The end vacuum must NOT fall below the permissible level of 10

-2

mbar / 0.0003 inHg.

Install the pump according to the permissible end vacuum or limit the end vacuum via a

vacuum controller.

Avoid the solvent accumulation in the vacuum module (option) as this would cause the vacuum

module to become an explosive area (Zone 0 or 1 or 2). The VDL vacuum drying oven located

on top of the module is not explosion-proof constructed in regards to its surroundings.