Fault description

Possible cause

Required measures

Heating (continued)

Chamber heating permanently,

set-point not held.

Controller defective.

Contact BINDER Service.

Pt 100 sensor defective.
Semiconductor relay defective
Controller not adjusted.

Calibrate and adjust controller.

Deviations from the indicated

heating-up times.

Oven fully loaded.

Charge the oven less or consider

longer heating-up times.

Deviations from the temperature

set-point in equilibrated state.

Invalid calibration

Use the delivered expansion racks

only. Do NOT change between

aluminum and stainless steel racks

Wrong temperature value

measured during calibration.

Reference temperature sensor

has insufficient contact to expan-

sion rack.

Fix the reference temperature sen-

sor with thermal conductive paste

or adhesive aluminum tape.

Leakage current when using a

thermo element not electrically

insulated

Mount a thermo element electrically

insulated from the rack.

Vacuum

Vacuum not held.

Door gasket defective.

Replace door gasket,

Safety glass panel defective.

Replace safety glass panel.

Gaskets of small flange connec-

tions (universal eccentric ring)

defective.

Replace gaskets of small flange

connections.

Inner tube connection leaky.

Contact BINDER Service.

Controller
No indication on the controller

display after turning on the ov-

en.

Oven not connected to com-

pressed air supply.

Connect the oven to the com-

pressed air supply before start-up.

Program doesn’t run correctly

after it is started.

Pressure > 125 ± 25 mbar / 3.69

±0.74 inHg).

A vacuum of 125 ± 25 mbar / 3.69

±0.74 inHg) or below is needed for

heating release. Only then start the

program.

Program duration longer than

programmed.

Inappropriate tolerances have

been programmed.

For rapid transition phases, do NOT

program tolerance limits in order to

permit maximum heating speed.

Program stops one section too

early.

Program line is incomplete.

When programming, define the end

value of the desired cycle by adding

an additional section with a section

time of at least one minute (with

setting “ramp”).

Programs have been deleted.

Change from 2 programs to 1

program or vice-versa

When changing, ensure that the

programs are no longer needed.

The controller returns to Normal

Display from any level.

No button was pressed for more

than 120 sec.

Repeat entries, enter the values

rapidly.

Message RANGE ERROR CH1

in Normal Display in Display 2

Sensor rupture between sensor

and controller

Contact BINDER Service.

Ramp temperature transitions

are only realized as steps.

Set-point programming type set

to “Step” in the User level (chap.

10).

Set the set-point programming type

to setting “Ramp” in the User level

(chap. 10).

Only qualified service personnel authorized by BINDER must perform repair.

Repaired units must comply with the BINDER quality standards.