VDL (E2.1) 04/2014

page 55/107

01

02

03

04

05

06

07

08

09

W

t

Figure 15: Possible temperature transitions

(with default setting “ramp” in the user level (chap. 10)

A heating up rate of 1.5 °C/min can be regarded as the realistic maximum.

If the oven is charged to full capacity, depending on the load, deviations from the specified

heating up times may occur.


Program entry as set-point ramp (example):

W/°C

t/min.

S01

S02

S03

S04

0

20

40

60

80

100

30

120

180

380

S05

Program table corresponding to the diagram (with default setting “Ramp”):

Section


SEC

Temperature

set-point

[ °C]

TEMP

Section length

[hh.mm]

TIME

S01

40

00:30

S02

60

01:30

S03

90

01:00

S04

90

03:20

S05

20

00:01

You can now enter the values of such a program table to the RD3 program controller (chap. 8.2).