11. Performance in case of failures

11.1 Performance after power failures

Power failure during fixed-value operation (Normal Display):
The entered parameters remain saved. After power returns, operation continues with the set parameters.
Power failure during program operation:
After the power returns, program course continues with the set-points that have been reached previously

during program operation.

11.2 Alarm messages

Alarm messages, e g. “RANGE ERROR CH1” in case of sensor rupture, are shown in Display 2 only in

Normal Display.
A buzzer can be activated / deactivated in the user level (chap. 10). It can be reset by pressing the “

EXIT

button. The alarm text shown in Normal Display goes off only if the cause of the alarm does not exist any

longer.

12. Safety device class 2 (DIN 12880)

The safety device protects the vacuum drying oven, its environment and the charging material from ex-

ceeding the maximum permissible temperature.
Please observe the guideline BGI/GUV-I 850-0 on safe working in laboratories (formerly BGR/GUV-R 120

or ZH 1/119 laboratory guidelines issued by the employers’ liability insurance association) and BGR 104

(formerly ZH 1/10 laboratory guidelines issued by the employers’ liability insurance association) (for Ger-

many).
Set the safety-device to the maximally permitted drying temperature according to the information panel

“temperature setting” (chap. 2.5), to prevent exceeding the maximum drying temperature in case of a

default. In the event of a fault in the temperature controller, the safety device (2)

permanently turns off

the vacuum drying oven. This status is reported visually by the indicator lamp (2a).
Check the operation of the safety device (2) by moving it slowly counter-clockwise until it is turned off. The

safety device cut-off is reported visually by the indicator lamp (2a).
Then release the safety device by pressing the reset button (2b), and turn on the VDL vacuum drying oven

as described previously.

Figure 16: Temperature safety device class 2

Function:
The safety device is functionally and electrically independent of the temperature control device and turns

off the unit at all poles.
If you turn the control knob (2) to its end-stop (position 10) the safety device will protect the appliance. If

you set it to a temperature a little above the set-point temperature, it will protect the charging material.

(2)