Start up, 1 settings at the rd3 program controller, Danger – BINDER VD 23 User Manual

Page 40

Advertising
background image

VDL (E2.1) 04/2014

page 40/107

5.

Start up

DANGER

Solvent-containing air penetrating and concentration in the electrical area of the

unit.
Explosion hazard.

Before turning on the unit via the key switch (1), sweep the area for electrical equipment

at least 15 minutes with compressed air or inert gas. Sweeping the area for electrical

equipment must take place during the entire operation, as otherwise the VDL vacuum

drying oven will shut off at all poles.

After connecting the supply lines (chap 4) and checking all safety instructions and conditions of installation

and safe operation, turn on the unit via the key switch (1):

Position 0: Oven is not operating
No consumption of compressed air or inert gas in the area for electrical equipment. If the fine dosage

valve (6) is open, inert gas may be used and can reach the ambient air.

Position I: Oven is operating

If no indication appears in the display of the program controller RD3, the compressed air sup-

ply is probably not properly connected.

Warming chambers may release odors in the first few days after commissioning. This is not a quality de-

fect. To reduce odors quickly we recommend heating up the chamber to its nominal temperature for one

day and in a well-ventilated location.

5.1 Settings at the RD3 program controller

After turning the unit on with the key switch (1) the controller is in Normal Display / Fixed value operation

mode.
Depending on the temperature value entered before, LED (7a) is lit if the heating is active, or no LED is lit

if the actual temperature is equal to or above the set-point.
In

Display 1 of the controller the actual temperature value is displayed.

With inactive week program timer:

In

Display 2 of the controller the actual date and time are displayed. Example:

15.05.06 13:52

With active week program timer:

In

Display 2 of the controller the actual date and time and the states of the week program timer chan-

nels are displayed. Examples:

15.05.06 13:52 - -

15.05.06 13:52 - �

15.05.06 13:52 � -

15.05.06 13:52 � �

Channel 1 Off,

Channel 2: Off

Channel 1: Off,

Channel 2: On

Channel 1: On,

Channel 2: Off

Channel 1 On,

Channel 2: On