15.5 Measuring access port vacuum 9 poles (option)

The measuring access port allows creating electrical connec-

tions for low voltages or sensors between the exterior and interi-

or of the VDL vacuum drying oven. A 9 poles plug for the outside

is included.

Figure 28: Measuring connection (15)

with measuring access port and supplied plug

Connections at the measuring access port

At the inner side of the measuring access port you can solder up to 9 cables. The inside connections

must be insulated against each other and against ground. Use 300 °C / 572°F solder.

The 9 contacts are conducted outside via the measuring access port. You can use them to connect a

device via the supplied plug.


Maximum load capacity of the switching contacts: 42 V AC/DC - 2A

DANGER

Electrical hazard.
Danger of death.
Damage to switching contacts and connection socket.

Do NOT exceed the maximum switching load of 42 AC/DC – 2 Amp.

Do NOT connect any devices with a higher load capacity.

Insulate the inside connections against each other and against ground. Use 300 °C /

572°F solder.

15.6 Communication software APT-COM™ 3 DataControlSystem (option)

The unit is regularly equipped with a serial interface RS 422 (9) that can connect the BINDER communica-

tion software APT-COM™ 3 DataControlSystem. The actual temperature value is given at adjustable in-

tervals. Programming can be performed graphically via PC. Up to 30 chambers with RS 422 interface can

be cross-linked. For further information, refer to the operating manual of the BINDER communication

software APT-COM™ 3.

Confirm that the interface mode is correctly set to “Modbus” in the user level (chap. 10).

Pin allocation of the RS 422 interface (9):

pin 2: RxD (+)
pin 3: TxD (+)
pin 4: RxD (-)
pin 5: TxD (-)
pin 7: GND