4 current inputs, 5 current clamps, 6 voltage inputs – CIRCUTOR AR6 Series User Manual

Page 14: Current, Inputs, Clamps, Voltage

Advertising
background image

AR6 portable network analyzer

Page 14 of 69

User Manual

4.4

CURRENT INPUTS

Inputs for current measurement

I1 I2 I3 IN Leakage

Input voltage

0…2 V

RMS

Measurement margin

from 1 to 120% of In

Primary current measurement In

Depends on clamp

Admissible overload

3 In

Consumption

≤0.0004 VA

Accuracy of the instrument
(excluding the clamp)

0.5%

± 1 digit (we must add the margin of error of

the clamp used to this number)

Special features

Earth leakage current measurement via low-
pass filter (option to activate or deactivate)

Input impedance

10 kΩ

4.5

CURRENT CLAMPS

The AR6 can use single-scale or multi-scale clamps. When using single-scale clamps the
instrument automatically recognises the clamps connected to it; in the case of multi-scale
clamps, when the clamps are connected a menu pops up to configure the current transformer’s
ratio (see current clamp configuration).

When using E-FLEX 54 cm multi-scale clamps the scales are as indicated in

Table 4-1

Table 4-1.- E-FLEX 54 cm clamps

Clamp

TURNDOWN

RATIO

LED

Range (2.5...100%)

E-FLEX 54 cm (See note 1)

200 A

blue

5…200 A

2,000 A

red

50...2,000 A

20,000 A

blue + red

500...20,000 A

Note 1: E-FLEX clamps have built-in LEDS to indicate the scale at which they are working.
When they work at a low scale the blue LED lights up, when they work at a medium scale the
red LED lights up and when they work at a high scale both LED light up.

4.6

VOLTAGE INPUTS

Inputs for voltage measurement

U1 U2 U3 UN U

EARTH

Input margin

Unom= 10 to 800 V

RMS

phase-neutral

Measured accuracy

0.5%

± 1 digit

Number of inputs

5

Peak voltage

2,500 Vpeak

Bandwidth

3.2 kHz

Input impedance

10 MΩ