TE Technology TC-720 User Manual

Page 76

Advertising
background image

C) Read the actual temperature of the control thermistor.

1. Let us assume the actual temperature is 2.50 °C.

2. The control command, CC, for “INPUT1” sensor temperature is 01.

3. There is no send value, so we calculate the checksum (SS) by adding the ASCII values of the characters

0, 1, 0, 0, 0, 0:

ASCII Character:

Hexadecimal Value:

0

30

1

31

0

30

0

30

0

30

0

30

Sum

121

16

The 8-bit checksum is the 8 least significant binary bits of the sum, represented as 21 in

hexadecimal.

4. Combining the characters in one string we send: (stx)01000021(etx).

5. If the temperature controller receives the command and the checksum is correct, it will send back:

(stx)00fa27(ack). The “27” at the end of the string is the checksum of “00fa”. The value 00fa

16

converts to 250

10

. This number, when divided by 100, is 2.50--the temperature in °C. As in the other

examples, if the controller did not calculate the correct checksum from the query it would send back:

(stx)XXXX60(ack).