TE Technology TC-720 User Manual

Page 83

Advertising
background image

43. FAN PWM FREQUENCY

Write Command: 3c

Read Command: 70

Interpret:

PWM frequency used for fan speed control (consult fan manufacturer for suitable frequency).

0 == 43,200 Hz

1 == 5,400 Hz

2 == 675 Hz

3 == 169 Hz

4 == 42 Hz

44. SET MODE

Write Command: 3d

Read Command: 71

Interpret:

0 == NORMAL SET

1 == RAMP/SOAK SET MODE

2 == PROPORTIONAL+DEAD BAND

45. RAMP/SOAK SEQUENCE START/STOP

Write Command: 08

Read Command: NA

Interpret:

Send 1 to start sequence

Send 0 to abort sequence

46. RAMP/SOAK status

Write Command: NA

Read Command: 09

Interpret:

Convert the returned hexadecimal value to binary and interpret the bits as follows:

All zeros means no sequence running.

Bit 0 == means sequence running.

Bit 1 == means SOAK stage.

Bit 2 == means RAMP stage.

For example, if a sequence is running during the ramp stage, the read command would return

0005

16

. This converts to 101

2

where Bit 0 (the rightmost bit) is set and Bit 2 is set.

47. SEQUENCE POINTER

Write Command: NA

Read Command: 0a

Interpret: Ramp/Soak sequence stage value 0 through 7

48. STAGE SOAK TEMPERATURES

Write Commands: a0 through a7

Read Commands: a8 through af

Interpret: To send a soak temperature, multiply the decimal value by 100

10

and convert to hexadecimal. To read

the soak temperature, convert the returned hexadecimal value to decimal, and then divide by 100

10

.

49. STAGE RAMP TIME COUNTS

Write Commands: b0 through b7

Read Commands: b8 through bf

Interpret: Ramp Time counts

83