To set the counter, To set the counter -71 – RIGOL DG1000Z Series User Manual

Page 121

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-71

To Set the Counter

You need to configure appropriate parameters for the counter.

1. Gate Time

Press Gate Time to set the gate time of the measurement system and the

default is “1.310ms”.

1ms

1.310ms

10ms

10.48ms

100ms

166.7ms

1s

1.342s

10s

10.73s

>10s

>10s

2. To Select the Parameters to be Measured

Press Select Meas to select the type of parameter to be measured by the

counter. The counter can measure the following parameters: frequency, period,

duty cycle, positive pulse width and negative pulse width. The default is

“frequency”.

3. Statistic Function

Press Statist to enable or disable statistic function. When the statistic function

is enabled, the instrument calculates the maximum, minimum, average and

standard deviation of the measurement values automatically and displays the

variation tendency of the measurement values in “Digital” or “Curve” mode.

1) To Select Display Mode

After the statistic function is enabled, press Display to set the display

mode of the statistic results to “Digital” or “Curve”, as shown in Figure 2-14

and Figure 2-15.

Advertising