RIGOL DG1000Z Series User Manual

Page 181

Advertising
background image

Chapter 5 Specifications

RIGOL

DG1000Z User’s Guide

5-7

Humidity Range

Less than 30℃: ≤95% Relative Humidity (RH)
30℃ to 40℃: ≤75% Relative Humidity (RH)
40℃ to 50℃: ≤45% Relative Humidity (RH)

Altitude

Operating: Less than 3000 meters

Non-Operating: Less than 15,000 meters

Mechanical
Dimensions

(W×H×D)

261.5mm×112mm×318.4mm

Weight

without package: 3.2kg

with package: 4.5kg

Interfaces

USB Host, USB Device, LAN

IP Protection

IP2X

Calibration

Interval

Recommend calibration interval is one year

Authentication Information

EMC

In line with

EN61326-1:2006

IEC 61000-3-2:2000 ±4.0kV (Contact Discharge)

±4.0kV (Air Discharge)

IEC 61000-4-3:2002

3V/m (80MHz to 1GHz)

3V/m (1.4GHz to 2GHz)

1V/m (2.0GHz to 2.7GHz)

IEC 61000-4-4:2004 1kV power lines

IEC 61000-4-5:2001

0.5kV (Phase to Neutral)

0.5kV (Phase to PE)

1kV (Neutral to PE)

IEC 61000-4-6:2003 3V, 0.15-80MHz

EC 61000-4-11:2004

Voltage dip:

0%UT during half cycle

0%UT during 1 cycle

70%UT during 25 cycle

Short interruption:

0%UT during 1 cycle

Electrical Safety

In line with

USA: UL 61010-1:2012,

Canada: CAN/CSA-C22.2 No. 61010- 1-2012

EN 61010-1:2010

Advertising