Seamless interconnection with oscilloscope, Seamless interconnection with oscilloscope -81 – RIGOL DG1000Z Series User Manual

Page 131

Advertising
background image

Chapter 2 Front Panel Operations

RIGOL

DG1000Z User’s Guide

2-81

Seamless Interconnection with Oscilloscope

DG1000Z could interconnect with

RIGOL oscilloscope which supports USB-TMC

seamlessly and rebuilt and output the waveform data collected by the oscilloscope

losslessly. The connection method between the instruments is as shown in the figure

below.

Operation Method:

1. Make sure that the measured signal is displayed steadily on the screen of DS1.

2. Connect the USB Host interface of DG1000Z with the USB Device interface of

DS1 using USB cable. Now, DS1 will enter “Remote” mode automatically.

3. Use DG1000Z to read the measured signal that is displayed currently on the

screen of DS1 (two methods):


 Press Store key at the front panel of DG1000Z to enter the store and recall

interface. At this time, the model identifier of DS1 will be displayed in the

“Directory Display Area”. Press the menu key File Type at the store and

recall interface of DG1000Z to select Arb File, and then set the Browser

menu of DG1000Z as “Dir” and use the knob of DG1000Z to move the focus

cursor onto the model identifier of DS1. Now, all channels and their on/off

status of DS1 will be displayed in the “File/Folder Display Area”. Set the

Browser menu of DG1000Z as “File”, and move the cursor onto the

channel to be read, and then press Read. At this point, DS1 will enter the

“Stop” mode automatically, and DG1000Z will read the arbitrary waveform

data (namely the waveform data collected by DS1) automatically. After the

read, the arbitrary waveform data is stored in the internal volatile memory

of the current channel of DG1000Z and DG1000Z switchs to arbitrary
waveform (Arb) mode (if the current mode is not arbitrary waveform mode)
automatically.


 Press Arb key at the front panel of DG1000Z to enter arbitrary waveform

mode, and press Select WformStored Wform to enter the store and

recall interface of arbitrary waveform (at this time, the backlight of the

Store key at the front panel of DG1000Z turns on). Set the Browser menu

of DG1000Z as “Dir” and use the knob of DG1000Z to move the focus cursor

onto the model identifier of DS1. Now, all channels and their on/off states

DG1000Z

DS1

USB Device USB Host OUTPUT INPUT

USB Cable BNC Cable

DS2

Advertising